QC6: Quantitative Test and Certification3

Ext. Barrel Module IFA030 - DAQ Run Number =1437

 

- DAMAGED FIBER: We have found some damaged fibers. Four fibers (cells D6 and D5) had a crack inside the fiber at the end glued into the aspirin tube . Another fiber (cell A16) had a little crack too. And we found two bad fibers (they gave us less signal) in cell B15 and B13. All af them were replaced using the teflon tube holes.

 

TileRow

Drop Before Repairing

Drop After Repairing

PMT 42 - Cell A16 - Fiber# 21

1

38.0%

<20%

PMT 42 - Cell A16 - Fiber# 21

3

36.5%

<20%

PMT 44 - Cell B15 - Fiber# 1

4

30.6%

<20%

PMT 24 - Cell B13 - Fiber# 17

4

25.1%

<20%

PMT 37 - Cell D_6 - Fiber# 3

8

26.0%

<20%

PMT 37 - Cell D_6 - Fiber# 3

10

31.1%

<20%

PMT 37 - Cell D_6 - Fiber# 13

8

35.2%

<20%

PMT 37 - Cell D_6 - Fiber# 13

10

36.6%

<20%

PMT 18- Cell D_5- Fiber#34

8

56.9%

<20%

PMT 18- Cell D_5- Fiber#34

10

60.4%

<20%

PMT 38- Cell D_6- Fiber#56

9

100.0%

<20%

PMT 38- Cell D_6- Fiber#56

11

100.0%

<20%

Files: raw_run1394.hbook - hist_run1394.ps - hist_run1437.ps - raw_data_run1437.ps - OPTICAL PROBLEMS RUN1394.txt - OPTICAL PROBLEMS RUN1437.txt