4
QC6: Quantitative Test and Certification3Ext. Barrel Module IFA002 - DAQ Run Number = 211
|
TileRow |
Drop Before Gluing |
Drop After Gluing |
|
|
PMT 37, Cell D6, Fiber #18 |
9 |
86.4% |
45.8% |
|
11 |
86.3% |
46.7% |
|
|
PMT 17, Cell D5, Fiber #53 |
9 |
85.5% |
36.0% |
|
11 |
86.6% |
47.2% |
|
|
PMT 18, Cell D5, Fiber #63 |
9 |
88.4% |
40.5% |
|
11 |
88.7% |
44.9% |
|
|
PMT 18, Cell D5, Fiber #65 |
9 |
73.8% |
32.5% |
|
11 |
75.4% |
26.4% |
|
PMT |
CELL |
TileRow |
Fiber # |
Drop |
Possible Problem |
|
22 |
A14 |
3 |
5 |
39.8% |
Coupling Related |
|
22 |
A14 |
3 |
12 |
62.9% |
Coupling Related |
|
PMT |
CELL |
TileRow |
Fiber # |
Drop |
Possible Problem |
|
29 |
A15 |
1 |
28 |
38.9% |
Coupling Related |
|
12 |
A13 |
1 |
25 |
27.2% |
Fiber Related |
|
12 |
A13 |
3 |
25 |
24.0% |
Fiber Related |
|
38 |
D6 |
9 |
23 |
25.5% |
Fiber Related |
|
38 |
D6 |
11 |
23 |
23.8% |
Fiber Related |
|
38 |
D6 |
10 |
8 |
27.8% |
Fiber Related |
|
PMT |
CELL |
TileRow |
Fiber # |
Drop |
Possible Problem |
|
42 |
A16 |
1 |
47 |
44.4% |
Fiber Related |
|
42 |
A16 |
3 |
47 |
42.2% |
Fiber Related |
|
29 |
A15 |
1 |
13 |
19.4% |
Fiber Related |
|
29 |
A15 |
3 |
13 |
29.9% |
Fiber Related |
|
41 |
A13 |
2 |
35 |
30.9% |
Fiber Related |
For Cell A16 (PMT 41, TileRow 2) the fiber #28 which had ~34% of Drop its profile was extracted and put it again obtaining a Drop ~30.9%, which indicates that the Optical problem could be related to the fiber itself.
Files:
raw_run211.hbook - hist_run13.ps - hist_run211.ps - raw_data_run211.ps - OPTICAL PROBLEMS RUN211.txt - OPTICAL PROBLEMS RUN13.txtJosep Flix (jflix@ifae.es)
Ilya Korolkov (korolkov@ifae.es)